TO DEVELOP HPLC METHOD FOR THE ASSAY OF MEMANTINE HYDROCHLORIDE TABLETS USING REFRACTIVE INDEX (RI) DETECTOR
Journal: Indo American Journal of Pharmaceutical Sciences (IAJPS) (Vol.04, No. 11)Publication Date: 2017-11-01
Authors : Tukaram B Sawant Dhananjay V Mane;
Page : 4391-4397
Keywords : Memantine hydrochloride; Assay; Liquid chromatography; Refractive Index detector.;
Abstract
Objective: To develop and validate a chromatographic method to determine the amount of drug (assay) in the tablets of memantine hydrochloride (MEM) using high performance liquid chromatography with refractive index (RI) detector. Methods: The chromatographic separation was achieved on C18 (250 × 4.5 mm, 5μ) column using isocratic mobile phase comprises of buffer (pH-6.0): Methanol (45:55 v/v) pumped at a flow rate of 1.0 ml/min. The detection of effluent was monitored using RI detector. Conclusion: The method was found to be simple, commercial, precise, accurate and robust which can be utilized for the determination of assay of MEM in tablets and capsules. Keywords: Memantine hydrochloride, Assay, Liquid chromatography, Refractive Index detector.
Other Latest Articles
- THE AGE DISTRIBUTION AND EFFECTIVENESS OF LOW DOSE TAMSULOSIN IN LOWER URETERIC STONES
- ISCHEMIC HEART DISEASE IN PATIENTS OF CHRONIC KIDNEY DISEASE ON MAINTENANCE HEMODIALYSIS
- FREQUENCY OF CRIMINAL ACTS / TYPE OF CRIME AMONG DIFFERENT PSYCHIATRIC DIAGNOSIS
- EFFECT OF ATROPINE AND GLYCOPYRROLATE IN PATIENTS WITH ORGANOPHOSPHATE POISONING
- DIFFERENTIAL DIAGNOSTICS OF SUBHEPATIC APPENDICITIS
Last modified: 2017-11-23 02:01:24