New Spectrophotometric Method Development and Validation of Lumefantrine

Journal: PharmaTutor (Vol.2, No. 5)

Publication Date:

Authors : ; ; ;

Page : 148-154

Keywords : Spectrophotometer; Lumefantrine; Validation; Antimalerial;

Source : Downloadexternal Find it from : Google Scholarexternal

Abstract

A Simple, sensitive, specific, spectrophotometric method has been developed for the detection of Lumefantrine in pure and Pharmaceutical formulations. The optimum condition for the analysis of the drug was established. Lumefantrine shows maximum absorption at 228 nm and obeyed beers law in the concentration range 10 to 50 ?g/ml. The correlation coefficient was found to be 0.999 and slope of line was found to be 0.0635. The percent S.D. for intra assay precision of the method was found to be 1.85% whereas Inter assay precision was found to be 0.44%. The sample solution was stable up to 24 hours. The assay results were found to be in good agreement with label claim. The proposed method was simple sensitive, precise, quick and useful for routine quality control.

Last modified: 2014-05-14 17:56:05