Biological Efficiency Testing According ISO 14698-1
Journal: Methods of Microbiology and Molecular Biology (Vol.1, No. 1)Publication Date: 2018-01-01
Authors : Roberto Ligugnana Roberto Montagna;
Page : 1-3
Keywords : Active sampling; Agar desiccation; Airborne particles; Airborne microbes; Air flow rate; Airborne microbial concentration; Agar medium; Bacterial cell; Biological agent; Bio sampler;
Abstract
The biological collection efficiency of the active multi-nozzle impactor air sampler TRIO. BAS MINI was compared to the efficiency of the slit to agar air sampler called CASELLA type that is used as traditional reference air sampler device. The TRIO.BAS biological efficiency was 82% vs Casella microbial air sampler [1,2]. Considering that the evaluation of biological efficiency between different microbial air samplers is always difficult and relative, the two results can be considered statistically comparable.