A Validated RP-HPLC Method for Estimation of Telmisartan and Metoprolol in its Bulk Form
Journal: International Journal of Biomedical Investigation (Vol.1, No. 2)Publication Date: 2018-07-01
Authors : Suthakaran R Bikshapathi DVRN Shankar CH Prem Kumar B; Ghouse S;
Page : 1-13
Keywords : Telmisartan; Metoprolol; RP-HPLC; ICH; Validation.;
Abstract
This method is used for the estimation of Metoprolol and Telmisartan present in dosage forms. Chromatogram was performed on Xbridge C18 (5u,4.6x250mm), with a mobile phase composed of methanol: water adjusted pH 3.5 with (58:42% v/v) at flow rate of 1mL/min. the identification was carried out at 224 nm. Parameters studied and reported as in ICH. The linearity range (60-140%) coefficients (0.999) and % recovery was found to be (98-102%).Retention time was (2.091) & (5.089). The test sample in all formulations were within in their label claims and this technique is used for analysis.
Other Latest Articles
- The Science of Dreams and Biochemistry of Midnight: A Questionnaire Study
- A New analytical method development and validation of Ambroxol, Montelukast and Levocetirizine in Tablet dosage form by RP-HPLC method
- RP HPLC Method Development and Validation for Quantitative Estimation of Tiotropium Bromide and Oldaterol
- Potential Link Between Proprotein Convertase Subtilisin/Kexin Type 9 and Alzheimer’s Disease
- Phytochemical Evaluation of Senna tora Linn
Last modified: 2018-12-30 14:57:02