Determination of trihalomethanes using gas chromatograph equipped with pulsed discharge electron capture detector (PDECD)
Journal: Asian Journal of Green Chemistry (Vol.4, No. 3)Publication Date: 2020-summe
Authors : Majid Hamouni Haghighat; Ali Mohammad Khah;
Page : 297-305
Keywords : Trihalomethanes; direct aqueous injection; Pulsed discharge electron capture detector Limit of detection;
Abstract
Trihalomethanes (THMs) are the result of the reaction between chlorine and organic matter in the chlorination process of water. Measurement of this compound in water is necessary due to the possible risks to human health. In this study, direct aqueous injection (DAI) using a gas chromatograph equipped with a pulsed discharge electron capture detector (PDECD) was used to analyze the THMs. The results showed that there is a significant linear relationship between concentration and peak area up to a concentration of 300 μg/L for THMs (chloroform, bromodichloromethane, dibromochloromethane and, bromoform). The limit of detection (LOD) was obtained 4.2, 4.0, 4.3 and 5.3 μg/L. Without any preconcentration of samples, small quantities of LOD values indicate the proper sensitivity of the detector and the analysis method. As a result, instead of the common type of electron capture detector (using 63Ni), pulsed discharge electron capture detector can be used.
Other Latest Articles
- Eco-friendly inhibitors for corrosion protection of metallic surfaces–a mini review
- A comprehensive DFT study on the adsorption of tetryl on the surface of graphene
- Green synthesis of titanium dioxide nanoparticles with Glycyrrhiza glabra and their photocatalytic activity
- Organocatalytic enantioselective one-pot synthesis of β-aminoketones via Mannich reaction
- A facile synthesis of benzimidazole derivatives over zinc sulfide nanoparticles as heterogeneous catalyst
Last modified: 2020-05-29 12:43:17