ResearchBib Share Your Research, Maximize Your Social Impacts
Sign for Notice Everyday Sign up >> Login

AN INNOVATIVE TRANSFORMER FOR ARTIFICIAL HEART SYSTEM

Journal: International Journal of Advanced Research in Engineering and Technology (IJARET) (Vol.11, No. 11)

Publication Date:

Authors : ;

Page : 2021-2028

Keywords : Transcutaneous transformer; Artificial heart; System; Transmission efficiency; Flux distribution;

Source : Downloadexternal Find it from : Google Scholarexternal

Abstract

A transcutaneous vitality transmission system was made out of a few loops, which shaped a transformer over the skin, a driving circuit, and an amending circuit. By utilizing coreless loops and a high driving recurrence (100–160 kHz), beyond what 25 W of electric power could be transmitted with 78.5% of greatest proficiency (dc to dc). In creature tests, the essential loop temperature during activity was under 39°C on thermograms. Following 10 months of implantation of an auxiliary loop covered with epoxy pitch, it was wrapped by a flimsy case of associating tissue. No conspicuous tissue response was perceived. Innovative geometrical construction of the transformer for an artificial heart system is explained in this article. The ferrite is used to construct the primary core and the amorphous magnetic material is used for constructing the secondary core of the transformer. 20 W is sufficient to drive the heat pump which can be easily transmitted has been experimentally verified. The gap in between the coils is generally of 5mm that provides the 90% transmission efficiency with 102.5 kHz switching frequency. Few characteristics of the constructed transformer are discussed and shown in this article.

Last modified: 2021-02-22 22:20:10