Ion Chromatography Method Development and Validation for Assay of Sodium Thiosulfate
Journal: International Journal of Science and Research (IJSR) (Vol.10, No. 4)Publication Date: 2021-04-05
Authors : Chanakya Thaker;
Page : 223-226
Keywords : Ion Chromatography; Suppressor; Conductivity;
Abstract
A new ion chromatography method was developed, validated and adapted for the assay of Sodium Thiosulfate Pentahydrate as active pharmaceutical ingredient. In this method, separation of Thiosulfate was done using IonPac AS19 column with eluent of 45mM Potassium Hydroxide using suppressed conductivity detection. Thiosulfate was eluted at around 5.3 minutes. Linearity rangeis 1.0 ? 100.0mg/L with regression coefficient value of 0.9999 for Thiosulfate. The Limit of Detection (LOD) value found was 0.15mg/L, and the Limit of Quantification (LOQ) value of 0.06mg/L. Validation parameters examined following suggestions of ICH are accurate ample for the supposed assay. The approach is confirmed as splendid method for assay of Sodium Thiosulfate as active pharmaceutical ingredient with excellent assay percentage values.
Other Latest Articles
- Comparative Studies of Prevalence of Ectoparasites of Cows and Goats Sold In Lokoja Area of Kogi State
- The Effectiveness of Posterior Femoral Glide Technique for Iliopsoas Pain and Tightness in Contemporary Dancers
- Radioclinical Spectrum in Ameloblastoma in 3 Patients
- An Overview of Fast Moving Consumer Goods (FMCG) of India
- The Theoretical Foundation of Medicine
Last modified: 2021-06-26 18:50:05