Analytical Method Validation Report for Quantitative Estimation of Diflubenzuron by HPLC Method
Journal: International Journal of Science and Research (IJSR) (Vol.9, No. 5)Publication Date: 2020-05-05
Authors : N. Sarath Babu; B. Ramachandra; N. Venkatasubba Naidu;
Page : 1443-1449
Keywords : Diflubenzuron; HPLC Method; Development and Validation;
Abstract
A simple, selective, precise and accurate high performance liquid chromatographic method for the analysis of Diflubenzuron in its formulations was developed and validated in the present study. The mobile phase consists of a mixture of acetonitrile, water and 1: 4 dioxane in the proportion 60: 40: 0.03 respectively (v/v/v/). This was found to give sharp peak of Diflubenzuron at a run time of 20 min. HPLC analysis of Diflubenzuron was carried out at a wave length of 260 nm with a flow rate of 1.3mL/min. The linear regression analysis data for the calibration curve showed a good linear relationship with a regression coefficient 0.999 in the concentration range of 50 % to 150 %. The linear regression equation was y= 1856x+304.2. The developed method was employed with a high degree of precision and accuracy for the analysis of Diflubenzuron. The method was validated for accuracy, precision, robustness, ruggedness and specificity. The precision, accuracy, sensitivity, short retention time and composition of the mobile phase indicated that this method is useful for the quantification of Diflubenzuron.
Other Latest Articles
- Wonderful Effect of Amrood (Psidium guajava Linn.) Drug of Unani System of Medicine
- Research on the External Effects of Chinas New Energy Vehicles
- Examining the Relationship between Learning Strategies and the Four Cs of Academic Resilience among School Students
- Pre-operative Prophylactic Tamsulosin for the Prevention of Postoperative Urinary Retention: A Single Centre, Prospective Case Control Study
- Utilization of Social Networking Sites: Association with Chemistry Academic Performance
Last modified: 2021-06-28 17:06:43