Spurious High Platelet Count by Automated Hematology Analyzer in Patient with Microcytic RBCS
Journal: International Journal of Science and Research (IJSR) (Vol.6, No. 1)Publication Date: 2017-01-05
Authors : Dr Rateesh Sareen; Dr Menka Kapil;
Page : 526-527
Keywords : Spurious platelet count; 5 part analyzer;
Abstract
We report a case of spurious elevation of platelet count on three consecutive occasions in a pediatric patient on six part Sysmex Hematology auto analyzer. Falsely low platelet counts are frequently reported in literature but falsely elevated Platelet counts are uncommon. The case emphasizes the important role of careful peripheral blood smear examination in case of high platelet counts. The blind folded reliance on automatic analyzers in hematology should be discouraged so as to provide best patient care.
Other Latest Articles
- Effect of Obesity on Psychological Profile of Adolescents
- Physiological studies of Helminthosporium Pathovars Isolated from Infected Maize Plant in the Region of Karnataka
- Correlation of Maxillary Sinus to the Roots of Maxillary Posterior Teeth and a Review of Literature
- Sustainable Employment Generation for Livelihood through MGNREGA Programme in Chikaballapur District of Karnataka
- Implications of Advanced Method
Last modified: 2021-06-30 17:35:27