Radiation Dose to Staff during Cardiac Interventional Procedures (Khartoum-Sudan)
Journal: International Journal of Science and Research (IJSR) (Vol.6, No. 11)Publication Date: 2017-11-05
Authors : Etidal Hussein. Khalafalla;
Page : 741-744
Keywords : Staff dose; interventional cardiology; Thermoluminescence dosimeters; radiation risks;
Abstract
The aim of this study was to measure and evaluate the staff radiation dose during cardiac procedures, such as coronary angiography (CA), percutaneous coronary interventions (PCI). Measurements were performed on Staff doses in (65) procedure using thermo-luminesces Dosimeters (TLDs) (LiF, Mg, Cu, P) at three anatomic locations (waist, chest and hand) for main operator and in the chest to assistant operator. While for patient, the dose measured using dose area product (DAP) meter. The radiation dose for the main operator's waist, chest and hand were 0.45, 0.47, 0.99mGy respectively and for the assistant operator's chest was 0.324mGy depending on TLD method indicating an increment by 28.8% relative to MPD. The monitoring of radiation workers is not established properly. It is obvious that high patient and staff exposure is due to the lack of experience and protective equipment's. The results presented are comparable with those published by other authors.
Other Latest Articles
- Background Analysis of the Golkar Party Conflict News Media (Critical Discourse Analysis)
- Self-Reference in Honorific of Sasak Speech Community: Sociolinguistics-Pragmatics
- Mechanical Vibration Analysis on a Testing Bench using Visualization Software
- Traditional vs Conventional Methods for the Management of Dandruff
- Impact of Illegal Gold Mining / Panning on Farming Activities in Zimbabwes New Resettlement Areas; A Case of Selected Farms along the Mazowe River in Bindura District; Mashonaland Central Province
Last modified: 2021-06-30 20:02:28