A simple, selective, linear, precise and accurate RP-HPLC method was developed and validated for rapid assay of 60 % SODIUM LACTATE NaC3H5O3. Isocratic elution at a flow rate of 1.2ml /min was employed on a symmetry phenomenex hypersil L17 (300mm x 7.80 mm, 8 m- rezex RHM - monosaccharide H+) column at ambient temperature. The mobile phase consisted of 0.1M Sulphuric acid. The UV detection wavelength was at 210 nm. Linearity was observed in concentration range of 0.001 - 0.005gm/ml. The retention time for 60 % Sodium Lactate was 4.0 min. The method was validated as per the ICH guidelines. The proposed method can be successfully applied for the estimation of 60 % Sodium Lactate.