Energy and Efficiency Calibrations for High Purity Germanium GEM30195 Coaxial Detector USING k0-IAEA Software
Journal: International Journal of Science and Research (IJSR) (Vol.4, No. 8)Publication Date: 2015-08-05
Authors : Joseph E.; Nasiru R.; Sadiq U.; Ahmed; Y. A.;
Page : 1055-1061
Keywords : Detector calibration; k0-IAEA software; NIRR-1; HPGe detector; FEPE;
Abstract
We have applied k0-IAEA software in this study to calibrate the high-purity coaxial detector (HPGe) GEM 30195 (ORTEC) at the Nigeria Research Reactor-1 (NIRR-1) laboratory, Centre for Energy Research and Training (CERT), Ahmadu Bello University, Zaria for use in the analysis of various samples. In order to achieve this, we first performed the energy calibration, plotted the energy calibration curves, and after that performed the full peak energy efficiency (FEPE) of the detector using two standard sources 137Cs and 152Eu at 15cm and 2cm geometries, and obtained values within the energy range of 121.78 1528.12 keV and 244.69 1408.00 keV respectively for 15 cm and 2 cm source-detector geometries with each recording the chi-square values of 1.0 and 3.6, where their z-values shows that all doublets at the photopeaks were correctly resolved. Besides, the results we obtained generally agrees with the limit of acceptance level of the software, and also shows good agreement with that carried out previously by other researchers.
Other Latest Articles
- Swarm Intelligence Algorithm with Guided Exploitations: A Case Study with Artificial Bee Colony Algorithm
- Sodium Nitrite Induced Biochemical Alterations in the Blood Serum and its Amelioration by Aqueous Extract of Libyan Propolis in Guinea Pigs
- An Assessment of the Effect of Feeding Graded Levels of Cassava Peel Meal Supplemented with Enzymes on Finisher Broiler Performance
- Efficient Contractor Selection and Bid Evaluation Methods for Construction Industry in India
- Role of Anthropology in Young Infant Feeding
Last modified: 2021-06-30 21:52:09