DEVELOPMENT AND VALIDATION OF RP HPLC METHOD FOR THE ESTIMATION OF SOFOSBUVIR IN BULK, TABLET AND INDUSTRIAL WASTE WATER

Journal: Indian Drugs (Vol.56, No. 06)

Publication Date:

Authors : ; ;

Page : 59-66

Keywords : *For Correspondence E-mail: vijayasree_2002@yahoo.co.in;

Source : Downloadexternal Find it from : Google Scholarexternal

Abstract

The study was aimed at development and validation of RP-HPLC method for estimation of sofosbuvir in bulk, pharmaceutical dosage form and pharmaceutical industrial waste. The chromatographic separation was performed on Agilent Syncronis C 18 (100 mm × 4.6 mm, 5μm) column, with a mobile phase comprising of a mixture of methanol: acetonitrile: water (45:30:25 v/v/V). The flow rate was 1.0 mL/min with detection at 260 nm. Retention time of sofosbuvir was found to be 2.040 min. As per ICH guidelines, the method was validated for linearity, accuracy, limit of detection, limit of quantitation, precision, robustness and system suitability. Linearity was found to be in the range of 4-24 μg/mL with regression equation y = 675284.x + 49120 and correlation coefficient 0.999. The low % RSD values indicate the method to be accurate and precise. The limit of detection (LOD) and limit of quantification (LOQ) were found to be 0.046 and 0.1400 μg/mL, respectively. The % recovery of tablets was found to be in the range of 99.9– 101.3%. It can be concluded that this validated HPLC method is easy, precise, accurate, sensitive and selective for estimation of sofosbuvir in bulk, tablet and applicable for analyses of in pharmaceutical industrial waste.