Comparison of optical biometry based on partial coherence laser interferometry (PCLI) principle to conventional applanation ultrasonic biometry
Journal: Indian Journal of Clinical and Experimental Ophthalmology (Vol.4, No. 3)Publication Date: 2018-09-01
Authors : Pooja H. V. Vijay Pai;
Page : 317-319
Keywords : 319;
Abstract
Purpose: Our aim was to compare the accuracy of intraocular lens (IOL) power calculation with two techniques namely optical biometry based on partial coherence laser interferometry (PCLI) principle and conventional ultrasound biometry. Materials and Methods: A comparative study was done in our hospital wherein 100 eyes of 100 patients undergoing phacoemulsification cataract surgery were recruited into it. Out of them, 50 eyes underwent optical biometry and 50 eyes underwent conventional applanation ultrasound biometry. Eyes were grouped according to axial length into long (>24.50), normal (22 to <24> Results: The MAE calculated was 0.60 +/- 0.40D and 0.20+/- 0.30D with the conventional applanation ultrasound and optical biometry respectively. It was found that the optical biometry gives more accurate values for eyeball with small and large axial length. Conclusion: The optical biometry parameters which uses the partial coherence laser interferometry (PCLI) principle happens to be the more precise method in IOL power calculation.
Other Latest Articles
- Strategic planning of a dental clinic in India
- THE METHOD OF ORGANIZATION OF PROJECT ACTIVITY OF YOUNGER SCHOOLBOYS IN THE PROCESS OF ENVIRONMENTAL EDUCATION
- VIDEO CLIPS AS AN ELEMENT OF DIGITALIZATION OF TEACHING OF INORGANIC CHEMISTRY OF THE FUTURE BACHELORS OF CHEMICAL EDUCATION
- Management and outcome of iatrogenic globe perforation during peribulbar block at tertiary eye centre
- Comparative evaluation of wound healing after ErCr:YSGG laser gingivectomy over conventional scalpel gingivectomy
Last modified: 2019-08-30 18:20:04